Clinical Trial: RDS MultiSense® SpO2 Calibration
Study Status: Completed
Recruit Status: Completed
Condition: Hypoxia
Study Type: Interventional
Official Title: RDS MultiSense® SpO2 Calibration
Brief Summary:
Data will be collected with the MultiSense® pulse oximetry system during non-motion conditions over the range of 70-100%. A Clinimark reference system, previously correlated to arterial blood CO-Oximetry will be the basis for comparison. A minimum of fifteen (15) up to fifty (50) healthy adult subjects, ranging in pigmentation from light to dark, with at least six (6) subjects with dark pigmentation (Fitzpatrick 5-6), will be enrolled in the study to meet the study design requirements defined by ISO 80601-2-61 and…